METERS
Temperature
PTM-9957, Temperature Controller/Monitor
PTM-9957
MRC.07.14
Features:
●
Thermocouple Temp. sensor: type
K/J/T/E/R/S.
●
Pt Temp. sensor: Pt 100 ohm.
●
Type K range: -100 to 1300°C, °C/°F.
Type J range: -100 to 1150°C, °C/°F.
Type T range: -100 to 400°C, °C/°F.
Type E range: -100 to 900°C, °C/°F.
Type R range: 0 to 1700°C, °C/°F.
Type S range: 0 to 1500°C, °C/°F.
●
Pt 100 ohm range: -199 to 850°C, °C/°F.
●
Resolution : 0.1 degree (< 1000 degree)
1 degree (
≥
1000 degree)
●
Control setting, Hi/Lo alarm setting.
●
Control relay output, alarm relay
output.
●
Control Relay will make action when
the reading value reach to control
value.
●
Alarm Relay will make action when the
reading value reach to high/low alarm
value.
●
Hysteresis value setting for control and
alarm function.
●
PoMicroprocessor circuit ensures high
accuracy and provides special
functions and features.
●
Power: 90 ACV to 264 ACV, 50/60 Hz.
●
RS-232/USB computer interface.
●
DIN size: 96 x 48 mm. Depth: 107 mm.
●
Optional type K probe: TP-01, TP-02A,
TP-03, TP-04, TP-05.
●
Optional Pt 100 ohm probe: TP-100B.
●
Optional USB cable, USB-01.
●
Optional date acquisition software,
SW-U801-WIN.
Model
PTM9957
Display
4 digits red LED, digit size : 14 mm.
Circuit
Custom chip of microprocessor LSI circuit.
Temp. Sensor type
Thermocouple Temp. sensor: Type K/J/T/E/R/S.
Pt Temp. sensor : Pt 100 ohm.
Sensor type can make the internal setting with default
Display units
°C/°F
Offset adjust
It can make the internal Offset setting with default.
Gain adjust
It can make the internal Gain setting with default
Sampling Time
Approx. 1 second.
Relay Output
Number: 2 relays
Relay 1: Control relay. Relay 2: High/Low alarm relay.
Max load: 0.5 ACA/250 ACV, 0.5 DCA/24 DCV
*Do not apply the relay contact load current > 0.5 A,
other wise the relay may be damaged permanently
without warranty.
Setting Function
1st layer setting procedures: CtLo (Control low limit),
CtHi (Control high limit), ALLo (Alarm low limit), ALHi
(Alarm high limit)
Second layer setting procedures: tyPE (Sensor type),
dEG (°C/°F), FiLt (Digital filter), CtHy (Control hysteresis
set), ALHy ( Alarm hysteresis set), oFSt (Offset
adjustment), GAin (Gain adjustment)
Data Output
RS 232 PC serial interface.
Operating Temp.
0 to 50°C.
Operating Humidity
Less than 80% R.H.
Power Supply
90 to 260 ACV, 50/60 Hz.
Power Consumption
Approx. 3.3 VA/AC 110V. Approx. 5.5 VA/AC 220V.
Weight
384 g
Dimension
DIN size: 96 x 48 mm, Depth: 110 mm.
Accessories
Included
Instruction manual.............................. 1 PC
Case holder with screw............................2 PCs
Optional
Accessories
Type K Temp. probe: TP-01, TP-02A, TP-03, TP-04, TP-05
Pt 100 ohm Temp. probe: TP-100B.
* Data Acquisition software, SW-U801-WIN.
* RS232 cable, UPCB-02.
* USB cable, USB-01.
* Real time SD card datalogger DL-9602SD
* GSM controller, GSM-889.
* Interface cable (cable between meter to GSM-889),
GMCB-89.
General Specifications:
Resolution
Range
Accuracy
0.1°C
-199.9 to 850.0°C
± (0.4 % + 1°C)
0.1°F
-327.0 to 999.9°F
± (0.4 % + 1.8°F)
1°F
1000 to 1562°F
± (0.4 % + 2°F)
Electrical Specifications:
PT 100 ohm
Sensor Type
Resolution
Range
Accuracy
Type K
0.1°C
-50.1 to -100.0°C -50.0 to 999.9°C ± (0.4 % + 1°C) ± (0.4 % + 0.5°C)
1°C
1000 to 1300°C
± (0.4 % + 1°C)
0.1°F
-58.1 to -148.0°F -58.0 to 999.9°F
± (0.4 % + 1.8°F) ± (0.4 % + 1°F)
1°F
1000 to 2372°F
± (0.4 % + 2°F)
Type J
0.1°C
-50.1 to -100.0°C -50.0 to 999.9°C ± (0.4 % + 1°C) ± (0.4 % + 0.5°C)
1°C
1000 to 1150°C
± (0.4 % + 1°C)
0.1°F
-58.1 to -148.0°F -58.0 to 999.9°F
± (0.4 % + 1.8°F) ± (0.4 % + 1°F)
1°F
1000 to 2102°F
± (0.4 % + 2°F)
Type T
0.1°C
-50.1 to -100.0°C -50.0 to 400.0°C ± (0.4 % + 1°C) ± (0.4 % + 0.5°C)
0.1°F
-58.1 to -148.0°F -58.0 to 752.0°F
± (0.4 % + 1.8°F) ± (0.4 % + 1°F)
Type E
0.1°C
-50.1 to -100.0°C -50.0 to 900.0°C ± (0.4 % + 1°C) ± (0.4 % + 0.5°C)
0.1°F
-58.1 to -148.0°F -58.0 to 999.9°F
± (0.4 % + 1.8°F) ± (0.4 % + 1°F)
1°F
1000 to 1652°F
± (0.4 % + 2°F)
Type R
1°C
0 to -1700°C
± (0.5 % + 3°C)
1°F
32 to 3092°F
± (0.5 % + 5°F)
Type S
1°C
0 to -1500°C
± (0.5 % + 3°C)
1°F
32 to 2732°F
± (0.5 % + 5°F)
Type K/J/T/E/R/S
275