Extra NewEra PDFs
|
SAFETY
|
TTL Inputs and Outputs - NE SyringePumpPro.pdf
|
DB9 on back of Panel . . . This is NOT an RS32 Port
|
Stall Detector on NE1000; NE1010
|
- A Safety Mechanism of LAST RESORT NOT to be used as a Flow Method Stop in LED Step programming or Via s’ware
- Program any STOP otherwise Pump Damage will Occur
|
Glass Syringe USE
|
ALWAYS Use a Safety Shield
|
NE-8000
|
- NO STALLl Detector Block is Installed
- Limit swirched on Guide Rails but always prgramm with these limis for Extra Saftey
- Sliding NUT Bloch ia solid all BRASS and can’t be manual adjusted use Pum pragraming to postion Syrine Start poine
|
NE-1000
|
- SS Syringes Not recommended Motor drive inadequate - Syringes are High Force to actuate
(Use NE8000 instea)
- Nylon 1/2 NutBlock prone to wear and skipping threads under lon term high forece
|
NE1000; NE1010; NE-8000
|
LAST RESORT SAFTEY Feature Motor has rorary vibration/sound sensor to STOP pump - expect wear on drive Shaft Thread, Motor belt; long term strain on motor windings and broker parts
NO Warranty Claims > DISCLAIMER Applies!
|
Force Limits
|
NE-1000 Max Flow 15lb , MIN Flow 35lb NE-1010 Max Flow 30lb , MIN Flow 70lb
NE-8000 Max Flow 100lb , MIN Flow 2000 > a Complete redesign NUt Block, Limit Switches, different motor belt, sturdier Motor NO STALL Block Device
|
Maintenance
|
Oil & Grease all so very lightly NE1000,> NE-1010 and >> NE8000 drive screw thread (and nut block thread) to reduce wear&tear
|
PPL Program XLS Sheets
|
Instructions PDF 
|
Standard PPLCreator Spreadsheet
|
XLS Files > ASK ! as these contain macro !
|
X and X2 upgrade PPL Creator Spreadsheet
|
XLS Files > ASK ! as these contain macro !
|
Peristaltic PPL Creator Spreadsheet
|
XLS Files > ASK ! as these contain macro !
|
Standard PPL Examples Spreadsheet
|
XLS Files > ASK ! as these contain macro !
|
PUMP Commands
|
see seperate PUMP Manuals
|